Tilt Series Extra Output Dialog

This dialog is used to control the saving of images to additional image buffers or files during a tilt series. The final Record or montage overview can be binned down and saved in buffers shown in a separate window. Four different types of images can be saved to files. First, the various tracking images taken by the Tilt Series Controller, and autofocus pictures, can be saved to files. Second, additional Record images can be saved at different defocus levels, or at different energy filter settings. These extra Record images are taken after the normal Record or montage is taken, and are saved to a separate file from the main tilt series to maintain the linkage between tilt and image number in the latter file. An entry will appear in the log for each image saved to an extra output file.  Third, in Low Dose mode a pair of Trial images can be saved, one from the other side of the Record area.  These images are taken like extra Record images, after the normal Record is taken, but they are not constrained to being at the same binning as regular Trial images.  Fourth, in STEM mode, images from other detectors than the primary one can be saved into a file.

When acquiring multiple tilt series through the Navigator, it is possible to have one extra file opened for each tilt series and closed at the end of the series.  Set up a script with the line
    SuffixForExtraFile  suffix
where 'suffix' will be added to the root of the filename of the regular tilt series output file.  For example, if the regular filename is 'series1.mrc' and the suffix is '_trial', the extra file will be named 'series1_trial.mrc'.  In the Navigator Acquire dialog, turn on the option to run a script before the main action and select the script with this command.  The extra file will be opened before the regular file, so set the file number for the extra output to 1 if there will be no other files open when the acquisition is started.  If there will be other files open, set the file number to one past the number of the highest file that will be open.

It is also possible to save more than one kind of extra output during multiple tilt series, but only one kind of output will go into separate files as just described; all of the series will write into the same file for the other kinds of output.

Stack Record/Overview images in separate window

Select this check box to display binned down Record or montage overview images in a window separate from the main window. This window will initially be opened at half the size of the binned images, and its title will be the name of the tilt series output file. When the window is the active one, hot keys can be used as usual to scroll through and zoom the display. (It may be necessary to click inside the window with the mouse to make it active.) Images will be labeled by their number in the buffer stack and by the Z value of the corresponding image in the output file.

This option can be turned on and off at any time during a series. If the window is closed, a new window will be opened when the next image is ready to be shown, but the previous images will be gone. At the end of tilt series, the window will no longer accept new images but it will be left open until you close it. If you start a new tilt series, a new window will be opened for its images.

Bin to N KB/image

This spin button lets you control how much the images are binned by adjusting the maximum memory allowed for each image. The memory requirement is the primary concern when keeping a stack that may contain over 100 images; for this reason, the displayed images are converted to bytes as well as binned down. The kilobytes per image and the maximum size of a square image are shown as you change this setting. The maximum size is limited by default to 512 by 512 pixels; this limit can be changed with a 'MaxStackWindowSizeXY' entry in the properties file.

View, Focus, Trial, Preview check boxes

These check boxes govern whether images of the particular type will be saved during the tilt series. You can change these selections at any point in the series.

File # and Name spin buttons

The spin buttons allow you to select a file number in which each kind of image will be saved. They can be set to file numbers up to the maximum number of image files allowed. If there is already an open file at a file number, its name will be shown; otherwise the selector will show 'Not open yet'. When the tilt series is started or resumed, the program will check the validity of the indicated file for each of the kinds of images being saved. Basically, each file must be compatible with images of the given type, based on the current camera parameters for taking that type of image. You can save different kinds of images into the same file as long as they have the same dimensions. You may not save any extra images into the main tilt series output file.

Montaged files cannot be selected for saving anything except extra Record images in TEM mode (i.e., a montaged file is not allowed for saving opposite Trials or STEM mode).  You will most likely want a montage that is the same size as the main output, but this is not required or done automatically by the program; you have to set up the montage of the needed size when you open the output file.

Extra Record/Trial radio buttons and text box

These buttons select whether one or more Record images are saved at different focus or filter settings or from different STEM detectors, or whether a pair of Trial images are taken. When a focus or filter series is selected, the text box is enabled and a series of entries can be made. Entries must be separated by spaces.

For a focus series, the values to enter are the change in focus from the current defocus target. Separate multiple values with spaces, e.g., '-2 -4 2'.

For a filter series, each entry consists of two values separated by a comma. The first value specifies the slit width: 0 to filter with the current width, -1 to take an image with the slit open, or a value greater than 0 to filter with the given width. The second value specifies the change in energy loss from the current value; positive and negative values are allowed. For example, if the regular Record image is set up to be zero-loss filtered with a slit width of 20, then entries of '-1,0 0,40 50,20' would result in three extra Record images: an unfiltered image, one with the regular slit width of 20 eV and an energy loss of 40 eV, and one with a slit width of 50 eV and an energy loss of 20 eV.

When you select the option to save images from multiple STEM detectors, the dialog will show a line listing the numbers of the detectors that will be acquired simultaneously with the main Record image, based on the current selection in the Record parameter set.  Only three channels can be acquired from simultaneously during a tilt series, i.e., two extra ones.  Separate acquisitions are needed to acquire from more than this number of detectors, or to acquire from ones that cannot be used simultaneously with the main Record detector.  Enter a space-separated list of additional channels to acquire in separate acquisitions in the text box.

Changes for Extra Records/Trials

The options in this group allow you to change the beam or the camera acquisition parameters for the extra Record or Trial images. Redundant methods are provided because there may be problems with one or two of them in any given case.

Keep file with bidirectional anchor images

Normally, the program stores the initial anchor image for a bidirectional tilt series in a file and deletes the file at the end of the tilt series.  If this option is selected, the image used to realign to the anchor is appended to the file and the file is not deleted.