SerialEM:
A Program for Automated Tilt Series Acquisition on Tecnai Microscopes Using Prediction
of Specimen Position
David
N. Mastronarde
Boulder
Laboratory for Three Dimensional Electron Microscopy of Cells, Dept. of
Molecular , Cellular, and Developmental Biology, University of Colorado,
Boulder, CO 80309
The
traditional approach to automated tilt series acquisition involves tracking
specimen position and measuring focus after every tilt, sometimes with
time-consuming magnification changes [1].
The predictable performance of modern goniometers has allowed the
development of a different method in which the translational specimen holder
movements are precalibrated with a rapid, coarse tilt series, then the full
tilt series is acquired relying on this calibration [2]. Although this method is efficient, with thick
specimens imaged at 300 KV in our laboratory we have found it susceptible to
drift, the less accurate adjustment of eucentric height for cryo-samples, and
other non-ideal features of the specimen or microscope. I thus developed a program, SerialEM, that
uses a new approach based on prediction of specimen position during the tilt
series from the position at previous tilts.
This method achieves both the speed of pre-calibration and the
reliability of the traditional approach.
The
new prediction method is illustrated in Fig 1.
At any given tilt angle, the program has information about the position
in X, Y, and Z at previous tilt angles.
A function is fit to each of these coordinates independently and the
position on the current tilt is predicted (Fig. 1, open circles). When there are only two data points and thus
no indication of the quality of the prediction (Tilt 3), the prediction is used
but the position is measured and corrected with a tracking image. With more data, the program examines both the
standard error of the new prediction (error bars), and the actual error of the
prediction on the previous tilt (vertical distance between open and filled
circles). If either of these errors is
too high (Tilt 4), the program will take a tracking image; but if both errors
are small enough (Tilt 5), the prediction is deemed reliable and no tracking
image is taken. The final recorded image
is used to measure and corrects position, to keep errors from accumulating. A similar procedure is used for focusing, except
here it is necessary to measure focus periodically.
Figure
2 shows results from a typical tilt series of a 300 nm thick plastic section,
with a tilt increment of 1º. The program
was able to skip tracking after 3 tilts, and focus only every 5º after 7
tilts. At -35º degrees, the program
reset the image shift to zero and moved the stage to compensate. (An alignment flaw limits the image shift on
the 300 KV Tecnai to less than 2 µm with small objective apertures.) Fig. 2 shows that after the image shift
reset, tracking and focusing were needed on every tilt for the next 6 tilts,
but thereafter tracking was unneeded and focusing was minimized. This tilt series took 45 minutes, of which at
least 30 were required simply to record the final 2K x 2K images from the Gatan
Megascan 795 camera. The total time is
thus comparable to that required for precalibration, as is the number of
pictures needed for focusing. In more
difficult situations where precalibration would fail, such as with extreme
drift, the series may take 10-20% longer as it tracks and focuses on a larger
fraction of tilts, but it still acquires an adequately aligned tilt series. The prediction method thus allows reliable
tilt series acquisition while minimizing collection time.
SerialEM
runs on FEI Tecnai microscopes with Gatan cameras. It presents an integrated user interface for
image acquisition, examination and storage. Many options for partially or fully automated
data collection make the program highly flexible; e.g., the user can stop the
tilt series if necessary, control the actions taken when it resumes, and even
conveniently back up to a previous tilt angle and replace the stored data from
that point onward. Tilt series can be
acquired in low-dose mode, with focusing and tracking images taken outside the
main recording area. SerialEM can also
automatically montage from multiple overlapping frames, making it useful for
cases where the area of interest is larger than the field of the camera. Finally, it incorporates control of a Gatan
Imaging Filter (GIF) and can acquire tilt series of energy filtered images.
References
[1]
A. J. Koster et al., Ultramicroscopy 46 (1992) 207-221.
[2]
U. Ziese et al., J. Microscopy 205 (2002) 187-200.
[3]
This work was supported by NIH/NCRR grant RR00592 and NIH Program Project grant
P01GM61306 to J. Richard McIntosh.
Figure
1. The prediction method. Y coordinate (perpendicular to the tilt axis)
is plotted versus tilt number for successive tilts. Closed circles are actual measurements; open
circles are predictions from previous positions.
Figure
2. The top three panels show changes in
X and Y coordinates and in focus setting (Z) during a tilt series. The lower panel shows a log of the exposures
taken during the tilt series: final unbinned images (vertical ticks), tracking
images binned by 4 (crosses), and focus measurements (circles). The latter involve taking three pictures to
measure beam-tilt dependent image shift.